NEET SS 2024 Diploma Radio Diagnosis Paper1 Question Paper with Solutions

Shivam Yadav's profile photo

Shivam Yadav

Updated on - Dec 10, 2025

The NEET SS 2024 Diploma Radio Diagnosis Paper1  Question  was conducted as part of the national-level super speciality exam for admissions into DM and MCh programmes. This paper primarily assesses advanced knowledge of pulmonary medicine, tuberculosis management, respiratory emergencies, infectious diseases, diagnostics, and clinical decision-making. Candidates were required to attempt 100 questions in 150 minutes, with a total of 400 marks. Each correct response carries 4 marks, while 1 mark is deducted for every incorrect answer following the NEET SS negative marking scheme.

NEET SS 2024 Diploma Radio Diagnosis Paper1  Question with Answer Key PDF

NEET SS 2024 Diploma Radio Diagnosis Paper1  Question Paper with Solutions PDF Download PDF Check Solutions
NEET SS 2024 Diploma Radio Diagnosis Paper1  Question with Answer Key PDF


Question 1:

a) What are the different types of MR angiography?

Correct Answer:
View Solution




Step 1: Understanding MR Angiography.

Magnetic Resonance Angiography (MRA) is a medical imaging technique used to visualize the blood vessels. The major types of MRA techniques include:

1. Time-of-Flight (TOF) MRA:
This technique is used for imaging blood vessels without the need for contrast agents. It relies on the movement of protons in flowing blood, which produces a strong signal that enhances the visibility of vessels, particularly arteries.

2. Phase Contrast (PC) MRA:
Phase contrast MRA uses the motion of blood to generate contrast. It can provide information about the direction and velocity of blood flow, making it particularly useful for assessing vascular conditions that affect blood flow dynamics, such as stenosis.

3. Contrast-Enhanced MRA:
In this method, a contrast agent is injected into the bloodstream, which allows for a clearer visualization of the blood vessels. It is commonly used for assessing more complex or deeper vascular structures.

Step 2: Differences between Types.

- TOF MRA is non-invasive and does not require contrast, making it ideal for arterial imaging.
- PC MRA is useful for flow measurements and gives insight into both flow velocity and direction.
- Contrast-Enhanced MRA is more effective for imaging complex or occluded vessels, but it involves the use of contrast agents, which could pose risks for patients with kidney issues.

Step 3: Conclusion.

Each MRA technique has its advantages and specific clinical applications depending on the need for either dynamic flow measurement or high-resolution imaging without contrast agents. Quick Tip: TOF is most suitable for arterial imaging, while PC MRA is best for evaluating blood flow. Contrast-enhanced MRA is effective for more complex vascular cases but requires contrast agents.


Question 1:

b) Describe the principle of time of flight (TOF) MR angiography, its advantages \& disadvantages and any four clinical applications.

Correct Answer:
View Solution




Step 1: Principle of Time-of-Flight (TOF) MRA.

TOF MR angiography is based on the principle of imaging moving blood. It works by exploiting the fact that blood, which has protons moving at a different velocity than the surrounding stationary tissue, can be distinguished due to its different magnetic resonance signals. Blood flowing into a slice receives a fresh magnetic pulse, enhancing the signal of moving blood, while stationary tissues do not exhibit this enhanced signal. This creates a high contrast image of arteries, especially in regions where blood vessels are located.

Step 2: Advantages of TOF MRA.

- Non-invasive: TOF MR angiography does not require the use of contrast agents, which is beneficial for patients with renal problems or those allergic to contrast media.
- High-resolution imaging: Provides clear, high-quality images of blood vessels, particularly in the brain and neck.
- No radiation exposure: Unlike X-ray angiography or CT angiography, TOF MRA does not expose the patient to ionizing radiation, making it a safer alternative for repeated imaging.
- Good for small vessels: Capable of imaging small and medium-sized arteries with high detail.

Step 3: Disadvantages of TOF MRA.

- Poor visualization of veins: Due to the low signal from venous blood, TOF MRA is less effective for imaging veins.
- Motion sensitivity: Patient movement, including respiratory or cardiac motion, can introduce artifacts into the images, reducing clarity and diagnostic value.
- Limited depth penetration: High field strength and long scan times are required for high-quality imaging, which can limit its applicability for deep or complex regions of the body.

Step 4: Clinical Applications of TOF MRA.

1. Intracranial aneurysms: TOF MRA is commonly used to visualize brain aneurysms due to its ability to create high-resolution images of cerebral arteries.
2. Arterial stenosis: It is effective in detecting stenosis (narrowing) of arteries, especially in the carotid and cerebral arteries, helping assess the risk of stroke.
3. Vascular malformations: TOF MRA can be used to detect arteriovenous malformations (AVMs) or other irregular vascular structures.
4. Preoperative evaluation of vascular anatomy: For surgeries like those involving arteriovenous fistulas, TOF MRA provides detailed information about blood vessels, aiding in surgical planning.

Step 5: Conclusion.

TOF MR angiography is a powerful tool for non-invasive, high-resolution imaging of blood vessels, particularly for arterial imaging, with various clinical applications, especially in neurology and vascular surgery. Its main limitation is poor vein visualization and sensitivity to motion artifacts. Quick Tip: TOF MRA is optimal for imaging arteries but may struggle with veins. It's particularly useful in neurological and vascular surgeries.


Question 2:

a) Discuss the harmful effects of radiation in humans.

Correct Answer:
View Solution




Step 1: Understanding the harmful effects of radiation.

Radiation, particularly ionizing radiation, can cause a variety of harmful effects in humans. This includes damage to cells and tissues, as well as increasing the risk of cancer. Ionizing radiation can strip electrons from atoms, creating free radicals that can damage DNA, proteins, and lipids within cells. These effects can lead to mutations, radiation burns, and in severe cases, death.

Step 2: Types of harmful effects.

- Acute effects: These occur with high doses of radiation over a short period. Symptoms can include nausea, vomiting, hair loss, and skin burns. In extreme cases, acute radiation sickness or even death can result.
- Chronic effects: These occur with low doses over a long period and can lead to cancers, genetic mutations, and organ dysfunction. Chronic exposure is particularly harmful as it may lead to diseases like leukemia, lung cancer, and thyroid cancer.

Step 3: Conclusion.

The harmful effects of radiation in humans are significant and can vary depending on the dose and duration of exposure. These effects emphasize the need for proper radiation safety measures. Quick Tip: Protecting against ionizing radiation is crucial for preventing long-term health effects, including cancer and genetic mutations.


Question 2:

b) What are the radiation protection measures that can be undertaken in a fluoroscopy unit?

Correct Answer:
View Solution




Step 1: Understanding radiation protection.

In a fluoroscopy unit, radiation protection measures are essential to safeguard both patients and healthcare workers from the harmful effects of radiation. These measures include the use of shielding, minimizing exposure, and employing safety protocols.

Step 2: Protective measures in fluoroscopy.

- Lead aprons and shields: Lead aprons are worn by patients and healthcare workers to absorb scattered radiation. Additional lead shields can be used to protect areas of the body not being imaged.
- Collimation: This technique involves limiting the size of the X-ray beam to the area of interest, thereby reducing the exposure to surrounding tissues.
- Distance: Increasing the distance between the radiation source and the individual reduces exposure. Fluoroscopy procedures should be conducted with as much distance as possible between the radiation source and the staff.
- Exposure time: Minimizing the exposure time by optimizing fluoroscopy settings and using pulsed fluoroscopy can significantly reduce radiation dose.
- Protective barriers: Fluoroscopy rooms are equipped with lead-lined walls to protect staff from scattered radiation.
- Personal protective equipment (PPE): For personnel, lead gloves, thyroid collars, and eye protection are important for minimizing exposure to scattered radiation.

Step 3: Conclusion.

Radiation protection in a fluoroscopy unit is crucial to ensure the safety of both patients and healthcare workers. The combination of shielding, distance, exposure time management, and proper equipment use helps minimize the risks associated with radiation. Quick Tip: Use of proper shielding, minimizing exposure time, and maintaining a safe distance are key strategies for radiation protection in fluoroscopy.


Question 3:

Enumerate the various MRI artefacts and discuss any four of them in brief.

Correct Answer:
View Solution




Step 1: Enumerating MRI artefacts.

MRI artefacts are distortions or errors in MRI images that are not related to the actual anatomy of the body. These artefacts can arise due to various reasons, including patient movement, equipment malfunction, or the intrinsic properties of the MRI scan. Common MRI artefacts include:


1. Motion artefacts

2. Ghosting artefacts

3. Metal artefacts

4. Chemical shift artefacts

5. Aliasing artefacts

6. Zipper artefacts

7. Truncation artefacts

8. Susceptibility artefacts


Step 2: Discussing four common MRI artefacts.


1. Motion Artefacts:
Motion artefacts are caused by patient movement during the MRI scan. These artefacts appear as blurring or doubling of the image. It is most common in brain, neck, and abdominal imaging. The artefact arises because MRI images are built up from a series of slices, and if the patient moves during scanning, the relative positions of the slices are altered.

2. Ghosting Artefacts:
Ghosting occurs when there is a mismatch between the time when the data is collected and the data being measured. This artefact typically results in a mirror image, often seen in phase encoding direction. It is caused by either patient movement or magnetic field inhomogeneity.

3. Metal Artefacts:
Metal artefacts arise when metallic objects (e.g., implants, dental work) distort the magnetic field, leading to signal loss or distortion in the images. These artefacts appear as dark or bright streaks around the metal, especially in high-field strength MRIs.

4. Chemical Shift Artefacts:
Chemical shift artefacts are caused by the difference in resonance frequencies of protons in different chemical environments. For example, fat and water have different frequencies, and when imaging, this difference can cause misplacement of fat and water signals, leading to artefacts at the interface between fat and water-containing tissues.

Step 3: Conclusion.

MRI artefacts are common in clinical practice and can often interfere with the diagnostic quality of the images. Understanding these artefacts and their causes is important for improving image quality and reducing distortion during scans. Quick Tip: To minimize MRI artefacts, ensure patient immobility, avoid metallic objects near the scanning area, and adjust scan parameters to reduce ghosting and aliasing effects.


Question 4:

What are ultrasound contrast agents? Discuss any four clinical applications of ultrasound contrast agents.

Correct Answer:
View Solution




Step 1: Understanding ultrasound contrast agents.

Ultrasound contrast agents (UCAs) are microbubbles that enhance the quality of ultrasound images by improving the visualization of blood flow and tissue structures. They consist of a gas core surrounded by a shell made from lipids, proteins, or polymers. UCAs are injected into the bloodstream and are used in contrast-enhanced ultrasound imaging to provide clearer images of the tissues, blood vessels, and organs. These agents reflect sound waves, creating a stronger echo that improves the resolution of the ultrasound images.


Step 2: Clinical applications of ultrasound contrast agents.

The following are four key clinical applications of ultrasound contrast agents:


1. Cardiac Imaging: UCAs are widely used in echocardiography to assess myocardial perfusion and identify areas of reduced blood flow in the heart. They help in the detection of heart conditions like coronary artery disease and myocardial infarction.


2. Liver Imaging: Ultrasound contrast agents are used to improve liver imaging, helping to detect liver lesions, tumors, and cirrhosis. UCAs enhance the visibility of abnormal growths and vascular structures in the liver, aiding in diagnosis and treatment planning.


3. Renal Imaging: UCAs are employed in renal imaging to evaluate kidney function and vascularity. They help in assessing conditions like kidney tumors, renal artery stenosis, and chronic kidney disease.


4. Tumor Imaging: UCAs are also used to enhance imaging of tumors in various organs. They provide better visualization of the tumor's blood supply and can help differentiate between benign and malignant tumors based on their vascular patterns.
Quick Tip: Ultrasound contrast agents are particularly useful in areas where conventional ultrasound imaging may lack sufficient contrast, such as soft tissues, blood vessels, and organs like the liver and kidneys.


Question 5:

a) Discuss the anatomy of the rotator cuff of the shoulder with well-labelled diagrams.

Correct Answer:
View Solution




Step 1: Understanding the rotator cuff anatomy.

The rotator cuff is a group of four muscles and their associated tendons that provide strength and stability to the shoulder joint. The rotator cuff allows for a wide range of motion of the shoulder while also maintaining the stability of the humeral head in the shallow glenoid cavity of the scapula. The four muscles of the rotator cuff are:


1. Supraspinatus: This muscle is located on the superior aspect of the scapula and is responsible for initiating the abduction of the arm.


2. Infraspinatus: Positioned below the supraspinatus, this muscle helps in lateral rotation of the shoulder.


3. Teres minor: This muscle assists in lateral rotation and adduction of the shoulder.


4. Subscapularis: Located on the anterior surface of the scapula, this muscle facilitates medial rotation of the arm.


Step 2: Diagram description.

A diagram of the rotator cuff should clearly depict the location and attachment of each muscle around the shoulder joint. Label the supraspinatus, infraspinatus, teres minor, and subscapularis. The diagram should also show the relationship between the muscles, tendons, and the humeral head.






Step 3: Importance of the rotator cuff.

The rotator cuff plays a critical role in shoulder movement and stability. It helps in lifting the arm, rotating the arm, and maintaining proper alignment of the humeral head in the glenoid cavity. Damage to any of these muscles or tendons can lead to pain, weakness, and limited range of motion, often referred to as "rotator cuff tear" or "impingement."
Quick Tip: In clinical practice, injuries to the rotator cuff are commonly seen in athletes or individuals involved in repetitive overhead activities, such as baseball players, swimmers, and manual laborers.


Question 5:

b) Discuss briefly the imaging features of various labral tears that can be seen in the shoulder joint.

Correct Answer:
View Solution




Step 1: Understanding labral tears.

The shoulder labrum is a ring of cartilage that surrounds the glenoid (the shallow cavity of the scapula) and helps to stabilize the shoulder joint. Labral tears can occur due to trauma, repetitive motion, or degenerative changes and are common in individuals who engage in overhead activities. Labral tears can be classified as either superior or inferior, depending on their location.


Step 2: Imaging techniques for labral tears.

Several imaging modalities are used to diagnose labral tears in the shoulder joint:


1. MRI (Magnetic Resonance Imaging): MRI is the most commonly used imaging technique to evaluate labral tears. The tear can appear as a disruption or a detachment of the labrum from the glenoid. MRI provides detailed soft tissue contrast and is highly sensitive for detecting labral tears, especially when combined with the use of contrast agents (arthrogram).


2. MR Arthrogram: In this procedure, a contrast dye is injected into the shoulder joint before performing the MRI. This enhances the visualization of the labrum, making it easier to detect tears or detachments. It is particularly useful in detecting small or subtle tears that may be missed on a standard MRI.


3. CT Arthrogram: Computed Tomography (CT) arthrograms may be used in cases where MRI is not conclusive or accessible. This technique involves injecting contrast into the shoulder joint and using CT scans to obtain detailed images of the labrum.


Step 3: Types of labral tears.

Labral tears can be categorized into different types:


1. SLAP (Superior Labrum Anterior to Posterior) Tears: These occur at the top (superior) part of the labrum, where the biceps tendon attaches. SLAP tears are commonly associated with overhead activities and can be seen on MRI or MR arthrogram as a detachment or tear of the superior labrum.


2. Bankart Lesions: These are tears of the anterior (front) part of the labrum and are typically seen in patients with shoulder dislocations. A Bankart lesion is best visualized on an MRI or CT arthrogram.


Step 4: Conclusion.

Labral tears are best diagnosed using MRI or MR arthrogram, with the choice of imaging modality depending on the clinical presentation and the severity of the tear.
Quick Tip: For accurate diagnosis of labral tears, consider using MR arthrogram when standard MRI does not provide sufficient detail, especially for detecting small tears or tears in difficult-to-visualize areas.


Question 6:

Discuss the following radiological procedures briefly:


a) Barium enema.

Correct Answer:
View Solution




Step 1: Understanding Barium Enema.

Barium enema is a radiological procedure used to examine the lower gastrointestinal tract, particularly the colon and rectum. In this procedure, a contrast medium containing barium sulfate is introduced into the colon through the rectum. The barium absorbs X-rays and appears white on the X-ray film, highlighting the inner surface of the bowel. Barium enema helps to detect abnormalities such as tumors, polyps, inflammation, or strictures in the colon. It is typically used for diagnosing conditions like colorectal cancer, Crohn's disease, and diverticulosis.


Step 2: Procedure.

The patient is asked to lie on an X-ray table while the contrast medium is inserted through a tube into the rectum. X-ray images are taken as the barium moves through the colon. In some cases, a double-contrast technique is used, where air is introduced after barium to provide clearer images of the mucosal lining.
Quick Tip: Barium enema is a valuable diagnostic tool, but it is now largely replaced by more advanced imaging techniques like CT colonography and colonoscopy.


Question 6:

b) Sonosalpingography.

Correct Answer:
View Solution




Step 1: Understanding Sonosalpingography.

Sonosalpingography (SSG) is a radiological procedure used to assess the fallopian tubes in women. It combines ultrasound imaging with saline or contrast medium to evaluate the patency (openness) of the fallopian tubes. The procedure is commonly performed in cases of infertility, where tubal blockage is suspected. SSG is a non-invasive and relatively simple procedure that helps identify conditions like tubal occlusion, adhesions, or structural abnormalities of the tubes.


Step 2: Procedure.

In sonosalpingography, a sterile saline solution is introduced into the uterine cavity through a catheter. Ultrasound is then used to monitor the movement of the saline and assess whether it passes freely through the fallopian tubes. If the tubes are open, the saline will flow freely; if there is a blockage, the saline will not pass through, which is visible on the ultrasound images.
Quick Tip: Sonosalpingography is less invasive and has fewer complications than other procedures like hysterosalpingography (HSG) for assessing tubal patency.


Question 7:

Discuss the imaging anatomy of nose and paranasal sinuses on NCCT with well-labelled diagrams. Also discuss the common anatomical variations that may be seen on NCCT PNS.

Correct Answer:
View Solution




Step 1: Understanding NCCT Imaging of Nose and Paranasal Sinuses.

Non-contrast computed tomography (NCCT) of the paranasal sinuses (PNS) is a key imaging technique used to evaluate the anatomical structures of the nose and sinuses. NCCT provides high-resolution images and is especially useful for detecting diseases and abnormalities of the nasal cavity and surrounding paranasal sinuses. This imaging modality is commonly used to assess conditions such as sinusitis, nasal polyps, and other sinus pathologies. NCCT is also essential for evaluating anatomical variations and pre-surgical planning.


Step 2: Anatomical Structures on NCCT PNS.

The following key anatomical structures can be clearly visualized on NCCT of the PNS:


1. Nasal Cavity: The nasal cavity appears as a midline structure and is bordered by the nasal septum and the lateral nasal walls. The mucosal lining of the nasal cavity is thin and homogenous. The nasal turbinates, which include the inferior, middle, and superior turbinates, are seen clearly.


2. Paranasal Sinuses: The paranasal sinuses, including the maxillary, frontal, ethmoidal, and sphenoidal sinuses, are well-delineated. The maxillary sinuses are typically the largest and are located lateral to the nasal cavity. The frontal sinuses are located above the eyes and are seen as air-filled cavities. The ethmoidal sinuses are situated between the eyes and are divided into anterior and posterior groups. The sphenoidal sinuses are located behind the ethmoid sinuses, near the central portion of the skull base.


3. Nasal Septum and Bony Structures: The nasal septum, composed of cartilage and bone, divides the nasal cavity into two halves. The bony structures of the nasal cavity, such as the maxillary bones, frontal bones, and ethmoidal bony plates, are also clearly visible.


Step 3: Diagram Description.

A well-labeled diagram should show the nasal cavity, the paranasal sinuses, and the surrounding bony structures. Label the nasal septum, turbinates, and each sinus (maxillary, frontal, ethmoid, and sphenoid). The diagram should highlight key features like the sinus cavities, nasal passages, and mucosal lining.






Step 4: Common Anatomical Variations on NCCT PNS.


Step 1: Understanding Common Variations.

While the NCCT scan provides detailed images of normal anatomical structures, it also helps identify common anatomical variations in the nose and paranasal sinuses. These variations may not be pathological but can have clinical significance, particularly in surgical planning.


Step 2: Common Anatomical Variations.

The following are common anatomical variations seen on NCCT PNS:


1. Deviated Nasal Septum (DNS): A deviation of the nasal septum is a frequent variation, where the septum is shifted to one side. This can cause airflow obstruction or contribute to conditions like sinusitis. The deviation is typically visible on NCCT as a shift in the midline of the nasal cavity.


2. Concha Bullosa: This refers to the pneumatization (air-filled cavity) of the middle turbinate. It appears as an air-filled structure within the middle turbinate on NCCT and can sometimes cause obstruction of the sinus drainage pathways.


3. Haller’s Cell: Haller’s cells are ethmoidal air cells located beneath the orbit. They may cause obstruction of the maxillary sinus ostium and contribute to sinusitis. On NCCT, they are seen as small, rounded air pockets near the lower part of the orbit.


4. Agger Nasi Cell: This is an anterior ethmoidal cell located near the frontal sinus. It can sometimes extend into the frontal sinus or cause obstruction of the sinus drainage pathways. These cells appear as small, well-defined air pockets on NCCT.


Step 3: Conclusion.

Common anatomical variations on NCCT PNS do not always require treatment but should be taken into account when diagnosing conditions or planning surgical procedures. Understanding these variations helps avoid misinterpretations of the scan.
Quick Tip: NCCT is often preferred over MRI in evaluating the sinuses because it is more effective in visualizing bone structures and detecting calcifications.
Awareness of anatomical variations such as DNS and concha bullosa is essential for accurate interpretation of NCCT PNS scans, especially in pre-surgical assessments.


Question 8:

What is the X-ray spectrum of an X-ray tube/target and how is it produced? What are the factors that affect the X-ray spectrum?

Correct Answer:
View Solution




Step 1: Understanding the X-ray Spectrum.

The X-ray spectrum produced by an X-ray tube consists of two components: the characteristic X-rays and the continuous X-rays (Bremsstrahlung radiation). The X-ray spectrum refers to the range of energies (frequencies) of X-ray photons emitted when electrons from a cathode in an X-ray tube are accelerated toward a target anode, typically made of tungsten. The electrons interact with the atoms in the target and produce X-rays.


Step 2: Production of the X-ray Spectrum.

- Continuous Spectrum (Bremsstrahlung Radiation): When high-energy electrons from the cathode hit the target, they interact with the nuclei of the target atoms. This interaction causes the electrons to decelerate rapidly, emitting energy in the form of X-rays. The energy of these X-rays can vary over a wide range, producing a continuous spectrum. The intensity of this spectrum increases with higher tube currents and higher electron energies.

- Characteristic Spectrum: When the electrons from the cathode interact with the electrons in the inner shells of the target atoms, they can eject the electrons from their orbits, leaving vacancies. Higher-energy electrons from the outer shells fall into these vacancies, emitting X-rays with specific energy levels. These X-rays are called characteristic X-rays because their energies correspond to the differences between the energy levels of the atomic shells.


Step 3: Factors Affecting the X-ray Spectrum.

Several factors can affect the X-ray spectrum produced by the X-ray tube:


1. Tube Voltage (kVp): The tube voltage or kilovolt peak (kVp) controls the energy of the electrons accelerated towards the target. A higher kVp results in higher energy X-rays, shifting the spectrum towards higher photon energies.


2. Tube Current (mA): The tube current (measured in milliamperes, mA) controls the number of electrons flowing from the cathode to the anode. An increase in tube current results in more X-ray photons being produced, leading to a higher intensity of the X-ray spectrum. However, this does not affect the energy of the X-rays.


3. Target Material: The target material, often tungsten, affects both the intensity and the quality of the X-ray spectrum. Tungsten has a high atomic number, which produces a higher energy and more efficient X-ray production. Other materials like molybdenum or rhodium may be used for specific applications (e.g., mammography).


4. Filtration: X-ray tubes are typically equipped with filters that remove low-energy (soft) X-rays from the spectrum. These filters improve the quality of the X-ray beam by reducing patient exposure to unnecessary low-energy radiation.


5. Anode Rotation and Cooling: The anode's rotation helps to dissipate heat generated during the production of X-rays. A rotating anode allows for higher tube currents and longer exposure times, resulting in a more intense X-ray spectrum. The cooling rate of the anode also affects the efficiency of X-ray production.
Quick Tip: Increasing the tube voltage (kVp) improves the quality (energy) of the X-ray spectrum, while increasing the tube current (mA) enhances the quantity (intensity) of X-rays produced.


Question 9:

What does AERB stand for? What are the major roles of AERB? Discuss the AERB guidelines that need to be followed for installation of a new digital radiography (DR) X-ray machine in your department.

Correct Answer:
View Solution




Step 1: What does AERB stand for?

AERB stands for Atomic Energy Regulatory Board. It is the regulatory body under the Department of Atomic Energy, Government of India, that oversees and ensures the safety of the use of ionizing radiation in various fields such as medicine, industry, and research. AERB’s primary goal is to regulate the safety of radiation exposure to the public, workers, and the environment.


Step 2: Major roles of AERB.

The major roles of AERB include:


1. Safety Standards and Regulations: AERB develops and enforces safety standards for radiation protection, nuclear safety, and radiological safety. These standards apply to various fields, including medical, industrial, and research activities.


2. Licensing and Authorization: AERB grants licenses and approvals for the use of radiation sources and equipment, including X-ray machines, radiation therapy units, and nuclear reactors.


3. Monitoring and Inspections: The board regularly inspects radiation facilities to ensure compliance with safety regulations and takes corrective actions when necessary. This includes periodic inspections of hospitals, clinics, and industrial units using ionizing radiation.


4. Training and Awareness: AERB provides training programs to professionals working with radiation, ensuring that they are aware of safety measures and protocols.


5. Radiation Protection: AERB works towards minimizing radiation exposure to workers, patients, and the public by setting safety standards, recommending protective equipment, and providing radiation protection guidance.



Step 3: AERB guidelines for installation of a new digital radiography (DR) X-ray machine.


1. Regulatory Approval: Before installing a new DR X-ray machine, approval must be obtained from AERB. The manufacturer and the medical institution must submit the details of the machine, including technical specifications, radiation safety features, and proposed installation site to AERB for review.


2. Radiation Safety Assessment: A comprehensive radiation safety assessment must be conducted to determine the potential radiation exposure to staff, patients, and the environment. The facility should have proper shielding, including lead walls or barriers, to ensure that radiation leakage is minimized.


3. Equipment Calibration: The new DR X-ray machine should be calibrated as per AERB standards. This includes verifying the dose calibration, performance of the X-ray tube, and image quality assurance. AERB may conduct periodic checks to ensure that the machine remains in compliance with safety standards.


4. Staff Training: The hospital staff, including radiologists, technicians, and radiation protection officers, must undergo proper training in the safe use of the DR X-ray machine. This training includes understanding radiation safety protocols, emergency procedures, and routine maintenance.


5. Installation of Monitoring Devices: Proper radiation monitoring devices, such as dosimeters and radiation detectors, should be installed near the X-ray machine. These devices help in tracking radiation exposure and ensuring the safety of personnel and patients.


6. Safety Protocols: AERB guidelines require the establishment of safety protocols for the operation of the DR X-ray machine. These protocols include procedures for patient positioning, staff protection, and equipment maintenance.
Quick Tip: AERB plays a vital role in ensuring the safe use of radiation and preventing accidents related to radiation exposure in various sectors.
Compliance with AERB guidelines ensures the safe operation of medical radiation equipment, preventing unnecessary exposure to radiation and ensuring patient safety.


Question 10:

Briefly discuss the following:


a) "F form" in PC-PNDT Act.

Correct Answer:
View Solution




Step 1: Understanding the PC-PNDT Act.

The PC-PNDT Act stands for the Pre-Conception and Pre-Natal Diagnostic Techniques (Prohibition of Sex Selection) Act, which was passed in 1994 in India to regulate the use of prenatal diagnostic techniques. The Act prohibits sex selection before or after conception and regulates the use of prenatal diagnostic techniques.


Step 2: "F form" in the PC-PNDT Act.

The "F form" is a form used in the context of the PC-PNDT Act. It is a specific form that must be filled out and maintained by the person or institution providing pre-natal diagnostic services, particularly for ultrasound examinations. The F form records important details, including the identity of the pregnant woman, details of the ultrasound procedure, and the reasons for conducting the procedure. The form helps track the use of prenatal diagnostic services to prevent misuse, such as sex-selective abortions. The F form must be submitted to the appropriate authorities as part of the regulatory requirements under the Act.


Step 3: Purpose of the F form.

The primary purpose of the F form is to ensure transparency in the use of prenatal diagnostic technologies and to prevent the misuse of such technologies for sex determination. It serves as a mechanism for monitoring and regulating these practices to ensure compliance with the law.
Quick Tip: The F form plays a crucial role in ensuring that prenatal diagnostic techniques are used ethically and in compliance with the PC-PNDT Act.


Question 10:

b) "Characteristic curve" of a photographic film.

Correct Answer:
View Solution




Step 1: Understanding the Concept of a Characteristic Curve.

The characteristic curve, also known as the H&D curve (after its creators, Hurter and Driffield), is a graph used to describe the relationship between the exposure of a photographic film to light and the resulting density (darkness) on the developed film. It is an essential concept in the study of photographic materials and is used to evaluate the performance of photographic films.


Step 2: Plotting the Characteristic Curve.

The characteristic curve is plotted with the following axes:
- The x-axis represents the logarithm of the exposure (often in terms of light intensity or exposure time).
- The y-axis represents the optical density of the film, which is a measure of how dark the film becomes after development.


The curve typically has three distinct regions:
1. Toe: The lower part of the curve where slight increases in exposure result in very little change in density. This corresponds to the film's underexposed region.
2. Straight Line (Linear Region): This is the middle section of the curve where the exposure and density have a linear relationship. This region is ideal for normal exposures.
3. Shoulder: The upper part of the curve where the film becomes saturated, and further increases in exposure do not result in much increase in density. This corresponds to overexposure.

Step 3: Importance of the Characteristic Curve.

The characteristic curve helps in understanding the contrast, speed, and tonal range of a photographic film. It is used to compare different films and to predict how a film will behave under various lighting conditions.


Step 4: Application.

In practical applications, the characteristic curve of a film is used in film development and in optimizing exposure to ensure high-quality images. Photographers and radiologists rely on the characteristics of the curve to make appropriate adjustments in exposure settings to avoid underexposure or overexposure.
Quick Tip: Understanding the characteristic curve is essential for proper film exposure and achieving optimal image quality in both traditional photography and medical imaging.